System Shutdown, Startup, and Reset
Resetting the Data System
58 TSQ Series Hardware Manual Thermo Scientific
Resetting the Data System by Using the Windows Shutdown and Restart Procedure
If possible, use the Windows shutdown and restart procedure to shut down and restart the
data system so that Windows can properly close programs and save changes to files.
To reset the data system by using the Windows shutdown and restart procedure
1. Choose Start > Shut Down from the Windows taskbar. The Shut Down Windows dialog
box appears.
2. To start the Windows shutdown and restart procedure, select Restart and click OK.
3. Observe the Windows shutdown and restart procedure on the monitor. Press
CTRL+ALT+DELETE when you are prompted to do so.
4. To complete the shutdown and restart procedure, click OK or enter your password (if you
have one) in the Logon Information dialog box.
Resetting the Data System by Turning the Personal Computer Off Then On
If you are unable to reset the data system by using the Windows shutdown and restart
procedure, proceed as follows.
To reset the data system by turning the PC off and on
1. Press the Power button on the personal computer to turn the personal computer off.
2. After several seconds, press the Power button on the personal computer to turn the
personal computer on.
3. Observe the Windows XP startup procedure on the monitor and press
CTRL+ALT+DELETE when you are prompted to do so.
4. To complete the shutdown and restart procedure, click OK or enter your password in the
Logon Information dialog box.
5. When the shutdown and restart procedure has completed, choose Start > All Programs >
Thermo Instruments > TSQ > TSQ Tune to display the EZ Tune window.
Note The communications link between the data system and the mass spectrometer
should automatically reestablish after you reset the data system. When this occurs the
Communication LED on the front panel of the mass spectrometer illuminates yellow and
then green. If the system is unable to reestablish the communications link, press the
System Reset button on the power panel of the mass spectrometer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSQ Quantum Access MAX and is the answer not in the manual?

Thermo Scientific TSQ Quantum Access MAX Specifications

General IconGeneral
BrandThermo Scientific
ModelTSQ Quantum Access MAX
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals