EasyManuals Logo

Thermo Scientific TSQ Quantum Access MAX User Manual

Thermo Scientific TSQ Quantum Access MAX
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
6
Maintenance
Frequency of Cleaning
Thermo Scientific TSQ Series Hardware Manual 77
Frequency of Cleaning
The frequency of cleaning the components of the mass spectrometer depends on the types and
amounts of samples and solvents that are introduced into the instrument. In general, for a
given sample and ionization technique, the closer a mass spectrometer component is to the
source of the ions, the more rapidly it becomes dirty.
Clean the sample transfer line, sample tube, and API probe at the end of each operating
day to remove any residual salts from buffered mobile phases or other contamination that
might accumulate during normal operation. See “Flushing the Sample Transfer Line,
Sample Tube, and API Probe” on page 66.
Periodically remove and clean the ion transfer tube and ion sweep cone of the API source.
See “Maintaining the Ion Sweep Cone and Ion Transfer Tubeon page 79.
The S-lens and exit lens (on the TSQ Vantage) or the tube lens and skimmer (on the
TSQ Quantum Access, TSQ Quantum Access MAX, and TSQ Quantum Ultra) of the
Q00 ion optics become dirty at a slower rate than the API probe, ion sweep cone, and ion
transfer tube. See “Cleaning the S-Lens and Exit Lens (TSQ Vantage)” on page 87.
The rf lens of the Q00 ion optics becomes dirty at a rate significantly slower than the
API source, tube lens and skimmer, or S-lens and exit lens. See “Cleaning the Q00 Ion
Optics” on page 91.
Q0 ion optics Clean Q0 quadrupole and
lenses L11 and L12
b
As needed
a
N/A
Mass analyzer Clean Q1, Q2, and Q3
quadrupoles and lenses
b
As needed
a
N/A
Ion detection system Clean ion detection system
(electron multiplier and
conversion dynode)
b
Whenever the side cover
plate of the vacuum
manifold is removed
N/A
Cooling fans Clean fan filters Every 4 months page 101
Ion detection system Replace electron multiplier
assembly
b
If noise in spectrum is
excessive or proper electron
multiplier gain can not be
achieved
N/A
Electronic modules Replace electronic module
b
If electronic module fails N/A
PCBs Replace PCB
b
If PCB fails N/A
a
The frequency of cleaning the components of the mass spectrometer depends on the types and amounts of samples and solvents that are
introduced into the instrument. Cleaning of the Q0, Q1, Q2, and Q3 quadrupoles is rarely (if ever) required.
b
A Thermo Fisher Scientific service engineer should perform this maintenance procedure.
Table 5. Mass spectrometer maintenance procedures (Sheet 2 of 2)
Mass spectrometer component Procedure Frequency Procedure location

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Scientific TSQ Quantum Access MAX and is the answer not in the manual?

Thermo Scientific TSQ Quantum Access MAX Specifications

General IconGeneral
BrandThermo Scientific
ModelTSQ Quantum Access MAX
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals