2. Specifications
2.1. Calibrator (Source)
Function Range / Values Best 1 Year Specification
Voltage DC 0 to ± 1050V ± 15ppm of setting
Current DC 0 to ± 22A ± 80ppm of setting
Voltage AC 1mV to 1050V (10Hz to 1MHz, sine-wave) ± 300ppm of setting
10μA to 22A (20Hz to 1kHz, sine-wave)
Clamp Meter Adaptor x50 turn AC/DC Current up to 1100A (DC, 45 to 90Hz) ± 0.5%
Capacitance 1nF, 10nF, 100nF, 1μF, 10μF, 100μF (100V Max) ± 0.25%
Inductance
1mH, 1.9mH, 5mH, 10mH, 19mH, 50mH, 100mH, 190mH, 500mH, 1H,
± 0.1%
Decade Resistance 1Ω to 1GΩ (decade values) ± 20ppm of setting
Full Range Resistance 1Ω to 120MΩ (variable) ± 100ppm of setting
Conductance 1s to 1ns (fixed values, decade steps) ± 20ppm of setting
Thermocouple Simulation -270 to 1820°C (type J, K, R, T, S, B, E, N) ± 0.15°C
PT100 Simulation -180 to 850°C ± 0.07°C
Oscilloscope Calibration
Amplitude 6mV to 200V and 6mV to 2V 50Ω (Square-wave or DC) ± 0.05%
Frequency/Period
0.1Hz to 100MHz / 10ns to 10s (fixed values 1, 2, 5 sequence)
± 0.1ppm (0.1Hz to 10MHz / 100ns to 10s)
± 20ppm (20, 50, 100MHz / 50, 20, & 10ns)
3 frequencies: 100Hz, 1kHz, 10kHz, settable from 0 to 100%
Fast-Rise < 400ps. Bandwidth checking up to 400MHz –
Option 9769: Scope 2.2GHz
Sweep
100MHz to 2.2GHz levelled sine-wave (0.5, 1, 1.5V pk-pk)
Amplitude ± 1%
2.2. 6.5 Digit Multimeter (measure)
Function Range / Values Best 1 Year Specification
Voltage DC 0 to 1000V 35ppm of rdg + 6ppm of rng
Current DC 0 to 3A 500ppm of rdg + 50ppm of rng
Voltage AC 0 to 750V 0.06% of rdg + 0.04% of rng
Current AC 0 to 3A 0.1% of rdg + 0.04% of rng
Resistance 0 to 100MΩ 100ppm of rdg + 50ppm of rng
Frequency 3Hz to 300kHz 0.01% of rdg
Thermocouple -270 to 1800°C (Type J, K, R, T, S, B, E, N) ± 0.5°C
PT100 -180 to 850°C ± 0.08°C
Note: See extended specifications data sheet for full technical data.