8-6
GUIDANCE AND MANUFACTURER'S DECLARATION
ELECTROMAGNETIC IMMUNITY
The RC-5000 intended for use in the following electromagnetic environments. The customer and
the user of the RC-5000 is advised to guarantee its use in the following environments.
Electromagnetic environments-
guidance
Conductive RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Immunity test IEC 60601 Test
level
Compatibility
level
3 V rms
150kHz to
80MHz
3 V/m
80MHz to
2,5GHz
3 V
3 V/m
It is preferable that the portable type or
movable type RF communication equip-
ment should not be used for any part of the
RC-5000 including the cables, within the
recommended separation distance which is
calculated by the formula applicable for the
frequency of the transmission equipment
Recommended separation distance
d =1.2 P
d =1.2 P
80 MHz to 800 MHz
d =2.3 P 800 MHz to 2.5 GHz
where, "P" is the maximum output rating
for the transmitter of which is expressed in
watts (W) as specified by equipment manu-
facturer and "d" is the recommended sepa-
ration distance expressed in the unit of
meter(m).
The intensity of electromagnetic field
emitted from the fixed transmitter is deter-
mined by field examination value
"a"
obtained from the electromagnetic field,
which is preferred to be les than the com-
patibility level for each frequency range of
.
" b"
Interference may occur close to the equip-
ment identified with the following mark.
Remark 1: Higher frequency is applied at 80MHz and 800MHz.
Remark 2: The guidance is not always applicable for all conditions. Electromagnetic transmission
is influenced by absorption and reflection from structure, objects, and human body.
8.5 EMC COMPATIBILITY DECLARATION
Table 204