23. WARNING AND ERROR MESSAGES
153
Error: Read Build Info.
Error: Read JOG Setting
Error: Read OS Parameter
Error: Read RCOR
Error: Read sysflg
Error: Self check
Error: Write sysflg
Press [OK] to cancel the message. If this error message appears frequently, contact your local
dealer.
Incorrect password.
Input password does not match set password.
Input correct password.
Input device name !!
Bluetooth device name not input. Input device name and complete device registration.
Input over 3 letters !
The input password consists of less than 3 characters. Input a password at least 3 characters in
length.
Motor error EXXX
A problem has occurred with the motor drive and operation stops.
Power the instrument OFF then ON to correct the problem.
If this error message appears frequently, contact your local dealer.
Need base pt. obs
During REM measurement, the observation of the target was not completed normally.
Reset and sight the prism and perform measurement again.
Need 1st obs
During missing line measurement, the observation of the starting position was not completed
normally.
Sight the starting position accurately and press [MEAS] to perform the measurement again.
New password Diff.
During new password setting, the passwords input twice are different.
Input new password twice correctly.
No solution
The calculation of the instrument station coordinates during resection does not converge.
Access the results and if necessary, perform the observations again.
N-prism not supported !!
Auto Tracking cannot be performed in reflectorless mode.
Use the prism to carry out automatic sighting.
Out of range
During gradient % display, the display range (less than ± 1000%) has been exceeded.
During REM measurement, either the vertical angle has exceeded horizontal ±89° or the
measured distance is greater than 9999.999m.
Install the instrument station far from the target.