29. EXPLANATIONS
187
The 0 index of the vertical circle of your instrument is almost 100% accurate, but when it is necessary
to perform particularly high precision angle measurements, you can eliminate any inaccuracy of the 0
index as follows.
If the power is cut off, the vertical circle indexing is ineffective. Do it again every time the power is
turned on.
PROCEDURE
1. Select “Obs.condition” in <Configuration>. Set “V
manual” (vertical circle indexing method) to “Yes”.
"22.1 Observation Conditions"
<V manual 0 set> is displayed.
2. Carefully level the instrument.
3. Accurately sight a clear target with a distance of
about 30m in the horizontal direction with the
telescope in face left.
Press [OK]. The vertical angle V2 is displayed
under "Take F2".
4. Turn the upper part through 180° and clamp it.
Then set the telescope in the face right position
and accurately sight the same target.
Press [OK].
The vertical and horizontal angles are displayed.
This concludes the vertical circle indexing
procedure.
29.2 Manually Indexing the Vertical Circle by Face Left, Face Right
Measurement