TROUBLE SHOOTING
7 This software doesn't run correctly because
the ROM version of the power supply is not
compatible with this KB-50/CV-5000PC.
Please update the power supply software to
version ××.×× or later.
Update the version of the CV power supply unit to
the specified one.
8 Error! Couldn't get Head-ROM version.
This software may not run correctly.
The CV power supply unit is connected to the
measuring head, CV-5000, through cable.
Check whether this cable is connected correctly.
Update the version of the measuring head CV-
5000 to the newest one.
9 This software doesn't run correctly because
the ROM version of the CV-5000 is not com-
patible with this KB-50/CV-5000PC.
Please update the CV-5000 ROM software
to version ××.×× or later.
Update the version of the measuring head CV-
5000 to the specified one.
10 Chart/Mask : Communication ERROR. Try to set the chart/mask again.
Check whether the chart connection of KB-50 is
set correctly.
Check whether the chart connection of the CV
power supply unit is set correctly.
11 Initialize ERROR!!
Do you change the SETTINGS, and con-
tinue?
If you select 'NO', this application will be
EXIT now.
The contents of setting are not correct. Touch the
[Yes] button in the message box and perform one
of the following operations.
Check the contents of the setting changed last.
Touch the [Default] button on the setting display
to return the data to the initial set values.
12 Can't reset CV!!
Do you change the SETTINGS, and con-
tinue?
If you select 'NO', this application will be
EXIT now.
The CV power supply unit is connected to the mea-
suring head, CV-5000, through cable.
Check whether this cable is connected correctly.
Then, touch the [Yes] button in the message box to
close the setting display.
13 Can't initialize Comm Port for CV!!
Do you change the SETTINGS, and con-
tinue?
If you select 'NO', this application will be
EXIT now.
Turn on the switch of KB-50 again.
14 CV is not READY!!
Would you like to change the SETTINGS,
and continue?
If you select 'NO', this application will be
QUIT.
The CV power supply unit is connected to the
measuring head, CV-5000, through cable.
Check whether this cable is connected correctly.
Turn on the switch of KB-50 again.
15 ERROR: Language DLL is not correct ver-
sion.
The file, which should display the set language,
cannot be used in the KB-50 of the current version.
Update the language file to the newest version.
16 The font for ××× character is not fount on
this system.
The font, which should display the set language, is
not installed in the KB-50 system.
It is necessary to update the Operating System of
KB-50.
17 Fail to get ××× language information. Turn on the switch of KB-50 again.
If the trouble is not solved, it is necessary to
update the Operating System of KB-50.
Message Phenomena and remedial measure