76
33 Communication ERROR. A communication error has occurred against the
device connected through the serial method.
• The CV power supply unit is connected to the
device through cable. Check whether this cable
is connected correctly.
• Check whether the setting is correct for the serial
port on the CV power supply unit, which is con-
nected to the device.
34 Course data load ERROR.
Maybe, course is not registered.
Start Semi Auto mode.
When the course optometry is about to be exe-
cuted, reading the course file has failed.
Register the course when the course optometry
method is used.
35 ERROR! Invalid Handle Value. The memory of KB-50 may be damaged.
Ask the serviceman to repair it.
36 ERROR! Can't Open Volume List. The memory of KB-50 may be damaged.
Ask the serviceman to repair it.
37 Save settings parameters ERROR. The memory of KB-50 may be damaged.
Ask the serviceman to repair it.
38 Initial setting has not been completed yet.
Do you set up again?
If you select 'NO', this application will be
EXIT now.
Necessary setting is not done correctly.
Touch the [Yes] button and perform setting cor-
rectly.
39 Need to set the parameters to each chart. When the test name is not displayed on the course
list in the course optometry registration display, the
parameters for the test are not set. Set the param-
eters for the test.
40 Save Examination Data ERROR. The memory of KB-50 may be damaged.
Ask the serviceman to repair it.
41 Register at least 1 chart. No chart is registered in the chart page.
At least, register one chart.
42 Fixation lamp chart is necessary. Register the fixation lamp chart in the chart page.
43 Only PC-2 Chart can be registered. In the chart page to which you tried to register a
chart, only the PC-2 far point chart can be regis-
tered.
Register the near point chart in other chart page.
44 This chart has already been registered. In the chart page to which you tried to register a
chart, the same chart has already been registered.
You cannot register the same chart.
45 Panel registration is needed. You tried to register the chart, which is not regis-
tered in the panel registration chart page of PC-2,
in the chart page.
In PC-2, other charts except those registered in the
panel registration chart page cannot be used.
46 Save program course date ERROR. The memory of KB-50 may be damaged.
Ask the serviceman to repair it.
47 Can't initialize exam/course list!! Reading the examination list/course list has failed.
Update the KB-50 to the newest version.
48 Can't show the course list.
No course is registered.
The course list cannot be displayed because the
course optometry is not registered.
Register the course when the course optometry
method is used.
Message Phenomena and remedial measure