TROUBLE SHOOTING /
29 Communication ERROR
(Timeout CTS-ON Check: CV SEND).
Communication ERROR
(Timeout CTS-OFF Check: CV SEND).
An error has occurred in the data transmission to
the measuring head CV-5000.
The CV power supply unit is connected to KB-50
through cable. Check whether this cable is con-
nected correctly.
The CV power supply unit is connected to the
measuring head, CV-5000, through cable.
Check whether this cable is connected correctly.
If the trouble is not solved, update the softwares
of the KB-50, the measuring head CV-5000 and
the CV power supply unit to the newest version.
30 Communication ERROR (Timeout CTS-ON
Check: CV RECEIVE).
Communication ERROR (Timeout CTS-OFF
Check: CV RECEIVE).
An error has occurred in the data reception from
the measuring head CV-5000.
The CV power supply unit is connected to KB-50
through cable. Check whether this cable is con-
nected correctly.
The CV power supply unit is connected to the
measuring head, CV-5000, through cable.
Check whether this cable is connected correctly.
If the trouble is not solved, update the softwares
of the KB-50, the measuring head CV-5000 and
the CV power supply unit to the newest version.
31 Communication ERROR
(Timeout CTS-ON Check: MC SEND).
Communication ERROR
(Timeout CTS-OFF Check: MC SEND).
An error has occurred in the data transmission to
the visual acuity chart connected through the
online system.
The CV power supply unit is connected to KB-50
through cable. Check whether this cable is con-
nected correctly.
The CV power supply unit is connected to the
visual acuity chart through cable. Check whether
this cable is connected correctly.
Check whether the setting is correct for the serial
port on the CV power supply unit, which is con-
nected to the visual acuity chart.
If the trouble is not solved, update the softwares
of the KB-50 and the CV power supply unit to the
newest version.
32 Communication ERROR (Timeout CTS-ON
Check: CV THREAD SEND).
Communication ERROR (Timeout CTS-OFF
Check: CV THREAD SEND).
An error has occurred in the data transmission to
the visual acuity chart connected through the
online system.
The CV power supply unit is connected to KB-50
through cable. Check whether this cable is con-
nected correctly.
The CV power supply unit is connected to the
visual acuity chart through cable. Check whether
this cable is connected correctly.
Check whether the setting is correct for the serial
port on the CV power supply unit, which is con-
nected to the visual acuity chart.
If the trouble is not solved, update the softwares
of the KB-50 and the CV power supply unit to the
newest version.