EasyManua.ls Logo

Topcon KR-800A - Patients Environment

Topcon KR-800A
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
71
GENERAL INFORMATION ON USAGE AND MAINTENANCE
PATIENT'S ENVIRONMENT
When the patient or inspector may touch the devices (including the connecting devices) or when the patient or
inspector may touch the person that comes into contact with the devices (including the connecting devices),
the patient's environment is shown below.
In the patient's environment, use the device conforming to IEC60601-1. If you are compelled to use any
device not conforming to IEC60601-1, use an insulation transformer.
Note 1: Use the device conforming to IEC60950-1.
CAUTION
Don't connect an additional power strip or an extension cord to
the system.
Don't connect the device which is not recognized as one compo-
nent of the system.
Devices applicable to the
patient's environment
Barcode reader (Note 1)
Do not use the power strip in the
patient's environment. Connect
the power supply of the device
to the commercial power supply.
2.5m
1.5m1.5m
Radius 1.5m

Table of Contents

Related product manuals