52
SETTING FUNCTIONS ON SETUP SCREEN
DATA COMMUNICATION (COMM)
Comm contains settings related to data output with the external device.
Description Options Details Initial value
KRT
(Print
setting on
KRT
mode)
KRT print layout
D/mm
KRT data is printed as follows,
D (corneal refractive power)/mm (corneal curvature).
D/mm
mm/D
KRT data is printed as follows,
mm (corneal curvature)/D (corneal refractive power).
Print form of KRT result
ALL Printout all measurement values.
ALL
AVE Printout only average value.
KRT ave. -HV or R1R2
HV
Display average of KRT measurement results is set to
HV (horizontal/vertical).
R1R2
R1R2
Display average of KRT measurement results is set to
R1R2 (flat/steep meridian).
KRT data -HV or R1R2
HV KRT measurement result is printed in simple format.
R1R2
R1R2 KRT measurement result is printed in full format.
KRT average
OFF Do not print KRT average value.
ON
ON Print KRT average value.
KRT cylinder
OFF Do not print kerato-cylinder value and axial angle.
ON
ON Print kerato-cylinder value and axial angle.
Corneal diameter
OFF Do not print corneal diameter.
ON
ON Print corneal diameter.
Description Options Details Initial value
Output data format
REF Only REF data are output.
ALLKRT Only KRT data are output.
ALL All data are output.
Communication Format
OLD OLD TOPCON format
OLD
NEW NEW TOPCON format
STD1 TOPCON STD1 format
STD2 TOPCON STD2 format
STD4 TOPCON STD4 format
CM1 Custom specification
CM4 Custom specification
Use of Output port
OFF RS-232C port is disabled.
OFF
ON RS-232C port is enabled.
Baudrate setting
2400bps Baudrate value:2400bps
2400bps
9600bps Baudrate value:9600bps