5–4
To access the Stat Test function, press ¢4² at the Setup menu. The
gauge displays the Stat Test menu:
1. Take STAT Test
2. Review STAT Test
3. Print STAT Test
<ESC> to Exit
To select a menu option, press the corresponding numeric key
or press ¢ESC² to return to the Setup menu.
Taking a Stat Test
Choose an area free of other nuclear gauges or radioactive
sources. Place the gauge on the reference standard block as
shown in Figure 9 on page 4–18.
To take a new stat test, press ¢1² at the Stat Test menu shown
above. The gauge prompts the operator to place the gauge on
the reference standard block with the source rod in the SAFE
(shielded) position. Ensure that the gauge is properly positioned
and press the ¢ENTER/START² key.
The gauge displays the progress of the stat test (twenty 1-
minute counts). Upon completion of the counts, the gauge
displays the results of the density stat test, including an
indication of whether the test passed or failed:
Density Stat Test
Avg. Counts: ####
R: #.### (PASS)
<ENTER> for Moist.
Press ¢ENTER/START² to view the results of the moisture stat
test. Follow the prompts to scroll through the results.