Model 3440 5–7
To access the Drift Test function, press ¢5² at the Setup menu.
The gauge displays the Drift Test menu:
1. Take Drift Test
2. Review Drift Test
3. Print Drift Test
<ESC> to Exit
To select a menu option, press the corresponding numeric key
or press ¢ESC² to return to the Setup menu.
Taking a Drift Test
To take a new drift test, press ¢1² at the Drift Test menu. The
gauge prompts the operator to place the gauge on the reference
standard block with the source rod in the SAFE (shielded)
position. Ensure that the gauge is properly positioned and press
the ¢ENTER/START² key.
NOTE
If it has been less than three hours or more than eight
hours since the last stat test, the gauge displays an
error message. Press ¢ENTER/START² to return to the
Setup
menu.
The gauge displays the progress of the drift test (five 4-minute
counts). Upon completion of the counts, the gauge displays the
results of the density drift test, including an indication of
whether the test passed or failed:
Density Drift Test
Avg. Counts: ####
R: #.### (Pass)
<ENTER> for Moist.
Press ¢ENTER/START² to view the results of the moisture drift
test. Follow the prompts to scroll through the results.