78
As prompted, place the gauge on the measurement site and
press any key. The gauge takes a 240-second count, during
which it displays the progress of the count. Upon completion,
the gauge calculates and displays the moisture offset (k) as
shown:
K: ##.##
Save This Value
For Later Use?
f To enable and use this moisture offset without storing it for
later use, press ¢NO². The gauge enables the offset and
returns to the Ready screen.
f To enable the moisture offset and store it for later use, press
¢YES². The gauge displays:
Select Memory Cell:
1: #.## 2: #.##
3: #.## 4: #.##
Press # to Select
The gauge can store the value in one of four memory cells.
Storing a new value in a cell erases the old value. Use the
number keys to store the value. This will replace any stored
values that were previously associated with the memory cell.
The gauge stores the moisture offset, enables it for use, and
returns to the Ready screen.
Disable the Moisture Offset
To disable the moisture offset value, press ¢6² at the Moisture
Offset menu. The gauge disables the moisture offset and returns
to the Ready screen.
NOTE
The moisture offset is also disabled when the gauge
is turned off.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler RoadReader 3430 and is the answer not in the manual?

Troxler RoadReader 3430 Specifications

General IconGeneral
BrandTroxler
ModelRoadReader 3430
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals