Overview, Continued
C-3
Table C-2. Determining Discharge Factor
Discharge
Time
Discharge
Factor
Discharge
Time
Discharge
Factor
Discharge
Time
Discharge
Factor
Discharge
Time
Discharge
Factor
0.083 hr
5
min
3.85 2 hr 1.43 8 hr 1.10 14 hr 1.03
0.166 hr
10
min
2.78 3 hr 1.30 9 hr 1.09 16 hr 1.02
0.249 hr
15
min
2.27 4 hr 1.22 10 hr 1.08 18 hr 1.01
0.332 hr
20
min
2.08 5 hr 1.18 11 hr 1.06 20 hr 1.00
0.5 hr
30
min
1.96 6 hr 1.14 12 hr 1.05 48 hr 1.00
1 hr
60
min
1.67 7 hr 1.11 13 hr 1.04 60 hr 1.00
8. Standby Discharge Factor (SDF) = _________ Alarm Discharge Factor (ADF) =
_________
9. Minimum required battery capacity calculation for a final discharge voltage of 20.4V (to
account for battery aging, a 1.17 multiplier is included).
Minimum battery capacity in Ah =
1.17 x [(A x C x SDF) + (0.0167 x B x D x ADF)] = _______
10. Minimum required battery capacity calculation for a final discharge voltage: 23V.
1.32 × [(A × C × SDF) + (0.0167 × B × D × ADF)] = _______
11. If the battery size initially selected is smaller than the result of step 9 above, repeat the
calculations using a larger battery size.
Battery Selection Calculations, Continued
Battery Selection