Ceilometer CT25K
User’s Guide CT25K-U059en-2.1
30
1st (TOP)
LEVEL
2nd LEVEL 3rd LEVEL 4th LEVEL 5th LEVEL DESCRIPTION
SET OPER_MODE
COMP_MONIT Run internal crosstalk compensation monitor until ESC
CONTINUOUS Operation Mode: Continuous measurement mode
STANDBY Standby mode, no measurement unless commanded by START
(initiates one cycle)
OPTION SKY_COND ON CODE Activate sky condition option
OFF Deactivate sky condition option
HUMITTER ON Enable humitter option
OPTION HUMITTER OFF Disable humitter option
BLOWER ON Enable blower related status and warning information, needs reset
OFF Disable blower related status and warning information
PORT TIMEOUT 2...60 Timeout for automatic CLOSE of dialog mode. Value in minutes.
Default: 2 minutes
MAINTENANCE B300 Set maintenance port bit rate to 300 bits/s
B2400 Set maintenance port bit rate to 2400 bits/s
B4800 Set maintenance port bit rate to 4800 bits/s
B9600 Set maintenance port bit rate to 9600 bits/s
DATA MODEM BELL_103 300 bits/s modem mode
V21 300 bits/s modem mode
BELL_212A 1200 bits/s modem mode
V22 1200 bits/s modem mode
V42_NORMAL 1200-2400 bits/s modem mode, with error correction (V.22bis + V.42)
V42_BIS 1200-9600 bits/s modem mode, with error correction and data
compression (V.22bis + V.42bis)
OFF Switch modem off and use serial line communication
INTERFACE RS232 Use RS232 serial line
RS422 Use RS422 4-wire serial line
RS485_2W Use RS485 2-wire serial line
RS485_4W Use RS485 4-wire serial line