5. Press ENTER to confirm the selection. A display similar to the following appears:
If you are using a transmitter with the optional display, the message HMI41
adjustment mode is shown on the upper row of the display.
6. Complete the adjustment using one of the following options:
Adjustment Using HMI41 as Reference Meter (page 64)
Adjustment Using HMI41 as Terminal (page 66)
6.4.2.1 Adjustment Using HMI41 as Reference Meter
Complete the steps (connect the cable and select adjustment mode) listed in
Adjusting Oset and Gain with HMI41 (page 63) before continuing with this
adjustment method.
When you carry out an adjustment using HMI41 as a reference meter, the HMI41 humidity
probe is used to show the correct humidity value (be sure that the reference probe has
previously been adjusted).
To complete the adjustment:
1. Check the display. A display similar to the one shown below should be visible. The
numbers on the
first line indicate the transmitter reading, and the numbers on the second
line indicate the reference probe reading.
HMT120 User Guide M211244EN-C