Figure 25 Analog output type (current or voltage) selection in Insight
Receiving analog output settings automatically from probe
The transmitter can receive the analog output settings automatically from the connected
probe in the following cases:
1. The transmitter is a new unit and has not been configured yet.
2. The user has removed the analog output configuration from the transmitter.
After connecting a probe, it is recommended to always review the analog output
settings with the Insight PC software or Indigo80 handheld indicator.
6.1.2 Analog output configuration example
This example shows how to use Insight to configure analog output 1 of the transmitter to use
the following settings when a humidity and temperature probe is connected to the transmitter:
4 ... 20 mA current output
Relative humidity measurement scaled to 0 ... 100 %RH
Clipping limit 0 %, error limit 5 %, and error output 3.6 mA
1. Connect the transmitter to the Insight software.
2. Select
and open the Configure device menu.
3. Open the General menu and select Current output from the dropdown. Store your
selection with Save.
4. Return to the Configure device menu and select one of the 3 analog outputs.
5. Select 4 ... 20 mA from the Output mode dropdown.
6. Select RH from the Output parameter dropdown.
7. Type 0 in the Scale low end field.
8. Type 100 in the Scale high end field.
9. Type 3.6 in the Error output level field.
10. Type 0 in the Clipping limit field and 5 in the Error limit field.
11. Select Save to take your configuration into use.
Chapter 6 – Configuring analog outputs