7.3.2 Restoring factory adjustment for analog outputs
You can use the Indigo80 handheld indicator to restore the factory adjustment for the analog
outputs of the transmitter.
Always restore factory adjustment before entering a new adjustment. This
prevents any possible earlier adjustments having an eect on the new adjustment
you make.
1. Connect the transmitter to the indicator.
2. Open the main menu by pressing .
3. Select Devices. If you have more than one device connected to the indicator, make a
further selection between the devices.
In this case, select Indigo300.
4. Select Calibrate / Adjust.
5. Select the desired output, for example, Analog output 1 adjustment.
6. Scroll down and select Restore factory adjustment > OK.
7. Confirm by selecting OK.
8. Exit the menu by pressing
or return to the main menu by pressing .
More information
Connecting to Indigo80 handheld indicator (page 36)
7.3.3 Analog output adjustment
If the analog output appears to be inaccurate, you can test and adjust the output level in the
Devices > Calibrate / Adjust menu of the Indigo80 handheld indicator. Note that each of the 3
analog output channels is adjusted separately. To carry out the adjustment, you need a
multimeter (or an alternative method for reading the analog output level) to verify the
transmitter test output level.
During the adjustment, the output of the transmitter's analog output channels is forced to a
fixed current or voltage level. The adjustment is a 2-point adjustment with a low point test
output and a high point test output: the fixed test output levels are listed in Table 6 (page 50).
Table 7 Transmitter analog channel fixed test output levels
Current and voltage test output points Fixed test output level
Low point test output for current output 4 mA
Indigo300 User Guide M212849EN-A