Model ATRT-03 (v) Operating Instruction
78
24.0 Enable Computer Interface
The following steps will put the ATRT-03 under computer interface mode.
Table 24-1. Change ATRT-03/03A/03B Real Time Clock
STEP DESCRIPTION DISPLAY ACTION
1 Select “Set Up” menu
1.TEST XFMR 04/12/00
2.SETUP 13:06:00
3.CALCULATOR
4.DIAGNOSTIC
Press key 2
2 Select “Next Page” menu
1.ENTER XFMR ID
2.REVIEW RECORD
3.SAV/RES RECORD
4.NEXT PAGE
Press key 4
3 Select “Computer Control” menu
1.COMPUTER CONTROL
2.SET TIME
3.SET TEST VOLTAGE
Press key 1
4 ATRT-03/03A/03B is under
computer interface
COMPUTER ITF MODE
*** CAUTION ***
CABLES MAY HAVE VLTG
“STOP” TO ABORT