EasyManua.ls Logo

Waters 2414 - Connect the Valve Pack Drip Tray

Waters 2414
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
June 11, 2015, 715004754 Rev. A
Page 41
Required tools and materials
•Sharp knife
3/16-inch ID Tygon
®
tubing (from the startup kit)
To connect the drip tray:
1. If you are connecting drip tray components from the 2414 RI detector to an
ACQUITY Arc system, proceed to step 3.
2. Cut a length of Tygon tubing sufficient to reach between the drip tray and the
waste container.
3. Connect the tubing to the white plastic fitting located under the three valves on
the front of the detector (see the figure on page 39).
4. Insert the other end of the tubing into the waste container.
2.6.4 Connect the valve pack drip tray
A factory-installed drip tray lies beneath the three valves of the valve pack. Be sure to
direct the drip tray tube to a suitable waste container.
Required tools and materials
•Sharp knife
3/16-inch ID Tygon tubing (from the startup kit)
To connect the drip tray:
1. If you are connecting drip tray components from the 2414 RI detector to an
ACQUITY Arc system, proceed to step 3.
2. Cut a length of Tygon tubing sufficient to reach between the drip tray and the
waste container.
3. Connect the tubing to the white plastic fitting located under the oven of the
detector (see the figure on page 39).
4. Insert the other end of the tubing into the waste container.

Table of Contents