EasyManua.ls Logo

Waters 2690 - 2.4 Preparing the Sample Management System

Waters 2690
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Preparing the Separations Module for Operation 25
2.4 Preparing the Sample Management System
Figure 2-13 shows the steps in preparing the sample management system for operation.
Figure 2-13 Preparing the Sample Management System for Operation
2.4.1 Purging the Sample Management System
Purge the sample management system whenever you:
Prime the solvent management system
Change solvents
See bubbles in the syringe
Start using the Separations Module at the beginning of each day
To purge the sample management system:
1. In the Main screen, press the
Menu/Status
key to display the Status screen
(Figure 2-7
).
2. Enter the appropriate solvent composition in the Composition fields.
3. Press
Direct Functions
to display the Direct Functions menu (Figure 3-4
).
STOP
Attention:
To avoid precipitating salts in the Separations Module, use an intermediate
solvent such as water when you change from buffers to high-organic-content solvents.
Start
Adjust the
Seal Pack
Prime the Needle
Wash Pump
Separations
Module is Ready
to Operate
Load the
Carousels
Purge the Sample
Management
System

Table of Contents

Related product manuals