Making Automatic Runs 45
5. When you have selected all the methods you want to transfer, press the
Export
Now
screen key. The file transfer begins.
4.2 Making Automatic Runs Under Millennium Control
If your Separations Module is configured for Millennium Chromatography Manager control
(Millennium version 2.xx or Millennium
32
), you make an automatic run from Run Samples
(QuickSet in Millennium versions earlier than 3.20) in the Millennium software. When the
Separations Module is running under Millennium control, the word “Remote” appears in
the banner area.
Making a Run Using Millennium Software
To make an automatic run, perform the following steps at the Millennium workstation:
1. Create a chromatographic system that includes the Waters 2690 Separations
Module:
• For Millennium software version 2.1x, create a system that includes the 2690
solvent management system and the sample management system. Refer to the
Millennium documentation for information on creating a chromatographic
system.
• For Millennium
32
software, create a system that includes the Waters 2690
Separations Module. Refer to the
Millennium Online Help
for information on
creating a chromatographic system.
2. Create a Millennium instrument method that uses the system you created in step
1, above. Refer to the Millennium documentation for information on creating and/or
modifying an instrument method.
Note:
Any parameter values not explicitly set in the Millennium instrument method
automatically take on the values in the Default separation method in the Separations
Module.
You can edit the Default separation method in the Separations Module to set
parameter values appropriate for your applications.
3. Create a method set that uses the instrument method you created in step 2,
above. Refer to the Millennium documentation for information on creating and/or
modifying a method set.