EasyManua.ls Logo

Waters 2690 - Accessing Direct Control Functions

Waters 2690
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34 Performing a Direct Control Run
To access the direct control functions:
1. Press the
Menu/Status
key (if necessary) to display the Status screen (Figure 3-2
).
2. Press the
Direct Function
screen key. The Direct Functions menu appears
(Figure 3-4
). Table 3-2 describes each of the direct functions.
3. Select the function you want to specify (press the number beside the function, or
use the arrow keys), then press
Enter
. Follow the instructions for each function.
Figure 3-4 Direct Functions Menu
2410 and 410
•Cell
•Col 1
•Col 2
Indicates the current status of the following parameters for the
2410 or 410 detector (if installed):
Cell Temperature
External Temp 1
1
External Temp 2
*
1
Enabled if the optional column heater module is configured with the Waters 2410 or
410 detector.
Table 3-1 Status Screen Parameters
(Continued)
Parameter Description

Table of Contents

Related product manuals