Performing a Direct Control Run 35
Table 3-2 Direct Functions
Direct Control Function Description
Dry Prime Opens the fluidic path (from the selected solvent
reservoir to the prime/vent valve) to replace air with
solvent, then performs a prime.
Wet Prime Replaces solvent in the path from the reservoirs to V1
and then to waste. Use to change the solvent(s) in the
system.
Purge Injector Removes mobile phase from the sample loop and from
the syringe and fills it with fresh mobile phase. As an
option, performs a compression test.
Purge 2410/410 Reference
(if applicable)
Purges the Waters 2410 or 410 RI detector reference
cell.
Program Sparge (if
applicable)
Sets the sparge rate for each reservoir.
Equilibrate Delivers mobile phase at the current conditions for the
specified period of time. Equilibrates the temperature
in the sample compartment and column heater at
preset values.
Condition Column Delivers mobile phase using the gradient table
specified in the current separation method without
injecting a sample or running the Events table.
Inject Samples Injects a sample one or more times from the specified
vial(s) using the selected method.