EasyManua.ls Logo

Waters ACQUITY UPC2 - ACQUITY 30-CM Single-Zone Column Manager Specifications

Default Icon
110 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
November 16, 2016, 715004521 Rev. C
Page 100
B.7 ACQUITY 30-cm single-zone column manager
specifications
Column compartment(or
compartments) temperature range
4.0 to 90.0 °C, settable in 0.1 °C increments; two
independent heat/cool zones per module, up to six zones in
a stacked configuration.
Column compartment(or
compartments) temperature
accuracy
±0.5 °C
Column compartment precision ±0.1°C
Column compartment(or
compartments) temperature
stability
±0.3 °C
Solvent conditioning Active pre-heating as standard.
Column tracking eCord™ Technology Information Management tracks and
archives achiral Viridis
®
columns.
Table B–14: CM-30S performance specifications
Feature Specification
Column capacity Eight selectable columns when bypass and waste are
omitted. Up to 4.6 mm internal diameter (ID), up to 300 mm
in length with filter or guard column.
Fittings 124,106 kPa (1241 bar, 18,000 psi), low dispersion, with
reusable column inlet fittings.
Column compartment temperature
range
20.0 to 90.0 °C, settable in 0.1 °C increments.
Column compartment temperature
accuracy
±0.5 °C
Column compartment temperature
stability
±0.5 °C
Solvent conditioning Passive pre-heating
Column tracking eCord Technology Information Management tracks and
archives column usage history.
Table B–13: CM-A and CM-AUX performance specifications (continued)