Figure 2–2: Console system view
2.3 Priming and equilibrating the system
The CO
2
side of the pump primes itself when you start the system and the flow is not turned on.
Prime only the co-solvent side of the pump using 100 percent co-solvent, typically methanol. The
priming process runs solvent through the lines to flush any particulates.
2.3.1 Priming the system
Always prime the system under these conditions:
• After replacing the needle.
• When the system has been idle for a relatively long time, such as overnight.
• During system startup.
• When changing co-solvent.
Navigate the tree to view and configure
instrument settings, view interactive displays
and plots, set maintenance counters, and
view logs
Flow and lamp status
indicators
Control panels show current
instrument conditions