November 16, 2016, 715004521 Rev. C
Page 22
Where required, tighten system fittings according to the procedures set forth in the overview and
maintenance guides for each instrument. Note particularly that the techniques used for
retightening fittings differ from those used to install fittings for the first time.
2.1.2 Powering-on the system modules
Requirement: You must power-on the system modules in the following order.
1. The ACQUITY UPC
2
workstation.
2. For configurations having two CM-30S modules running Empower software, the secondary
column manager.
3. For configurations having two CM-30S modules running Empower software, the primary
column manager.
See also: “Setting up a primary and secondary column manager” for information explaining
how to configure a system that includes two column managers.
Requirement: Before you power-on the detector, set the flow on the ccBSM to fill the flow
cell with solvent, thus removing any air.
4. All remaining system modules in no particular order.
5. Observe the LED indicator on the right-hand side on each module to determine its
operational status.
Requirement: If communication with system modules is interrupted, you must reset the
software. To do so, right-click on a control panel, and select Reset or click Reset on the
control menu item in the console instrument page.
Table 2–1: ccManager status
LED display Module status
Steady green Liquid CO
2
is flowing.
Flashing green Setting the valve position and preparing for normal
operation.
Flashing red Error state. The system vents and the flow stops.
Refer to the console for the error message.
Steady red Serious system failure within the module itself,
possibly preventing further operation and
communication with the workstation computer.
Flashing red/green Request for visual indication of the module.
Table 2–2: ccBSM status
LED display Module status
Steady green Flowing and operating normally.
Flashing red Error state. The pump and solvent flow stops. Refer
to the console for the error message.
Steady red Serious system failure within the instrument module
itself, preventing further operation, and may prevent
communication with the PC.
Flashing red/green Request for visual indication of the instrument.