B Specifications
The specifications presented here depend on the conditions in individual laboratories. Refer
to the ACQUITY
®
UPC
System Site Preparation Guide, or contact the Waters
®
Tech nic al
Service organization for additional information about specifications.
B.1 System specifications
Table B–1: ACQUITY UPC
2
system specifications
Feature Specifications
Operating flow rate range 0.010 to 4.000 mL/min, in 0.001-mL increments.
Maximum operating pressure and
maximum flow rate
3.25 mL/min up to 41,369 kPa (413 bar, 6000 psi).
4 mL/min, to a pressure of 29,303 kPa (293 bar,
4250 psi).
ABPR Range: 10,000 to 400,000 kPa (100 to 400 bar, 1450 to
5801 psi)
Noise: < ± 0.5 bar
Unattended operation Leak sensors, full 96-hour diagnostic data display
through the console software
Liquid CO
2
Minimum purity level of 99.97%
Table B–2: Instrument control specifications
Feature Specification
External control
Empower
®
software or MassLynx
®
software
External communications Ethernet interfacing via RJ45 connection to host PC
Event inputs/outputs Rear panel contact closure, TTL inputs, outputs, or both