November 16, 2016, 715004521 Rev. C
Page 23
See also:
•The ACQUITY 30-cm Single-Zone Column Manager Overview and Maintenance
Guide for information about the CM-30S status indicators.
•The ACQUITY UPLC Column Manager - Active and Column Manager - Auxiliary
Overview and Maintenance Guide for more information about the CM-A status
indicators.
•The ACQUITY UPC2 PDA Detector Overview and Maintenance Guide for information
about the PDA detector’s status indicators.
2.2 Starting the console software
The console is a software interface through which you configure the system settings and
instrument parameters, monitor system performance, run diagnostic tests, and maintain the
system and its modules.
In the console, you navigate to visual representations of each module and their components.
You can start the console from within the data system that you are using.
See also: The Empower software or MassLynx software online Help for details about
starting and logging on to the data system. Refer also to the Waters Console online Help.
Table 2–3: SM-FL status
LED display Module status
Steady green Indicates the SM-FL is operating normally, attempting
to complete any outstanding samples or diagnostic
function requests. When sample and diagnostic
function requests are finished, the LED reverts to the
unlit mode.
Flashing red Indicates that an error stopped the module’s
operation. Refer to the console for information
regarding the error.
Steady red Indicates a module’s failure that prevents its further
operation. Cycle the module’s power. If the LED is still
steady red, contact your Waters service
representative.
Flashing red/green Request for visual indication of the module.