What to do if... Multi 3620 IDS
105 ba77169e01 04/2016
Data transmission to
USB printer does not
work
Error message,
Only use
rechargeable
batteries!
Ni-MH 1.2 V, >2000
mAh
Error message,
Memory error 1
15.2 pH
Error message
OFL, UFL
Cause Remedy
– The USB-B interface is con-
nected to a PC
– Disconnect the PC from the USB-B
interface
– Connected USB printer does not
print
– Use a suitable USB printer (see sec-
tion 13.4)
– Check the printer settings (see sec-
tion 13.4)
– No power pack connected – Connect the power pack
Cause Remedy
– A battery voltage untypical for
NiMH batteries was identified.
– Disconnect the power pack from the
meter.
– Check whether the batteries in the
battery compartment are suitable
(Ni-MH 1.2 V, >2000 mAh).
– Unsuitable batteries:
Insert suitable batteries and recon-
nect the power pack.
– Suitable batteries:
Connect the power pack and close
the error message with <[ESC]>
or <M>.
CAUTION
The charging of unsuitable battery types may cause damage.
Cause Remedy
– Instrument storage was not
found
– Please contact the service.
Cause Remedy
IDS-pH sensor:
– Measured value outside the
measuring range
– Use suitable IDS-pH sensor
– Air bubble in front of the junction – Remove air bubble