Multi 3620 IDS What to do if...
ba77169e01 04/2016 106
Error message,
Error
No stable measured
value
Air in the junction Extract air or moisten junction
Gel electrolyte dried out Exchange IDS-pH sensor
Cause Remedy
Cause Remedy
IDS-pH sensor:
The values determined for zero
point and slope of the IDS-pH
sensor are outside the allowed
limits.
–Recalibrate
Junction contaminated Clean the junction
IDS-pH sensor broken Exchange IDS-pH sensor
Buffer solutions:
The used buffer solutions do not
agree with the set buffer set
Set different buffer set
or
Use different buffer solutions
Buffer solutions too old Use only once.
Note the shelf life
Buffer solutions depleted Change solutions
Cause Remedy
IDS-pH sensor:
Junction contaminated Clean the junction
Membrane contaminated Clean membrane
Test sample:
pH value not stable Measure with air excluded if neces-
sary
Temperature not stable Adjust temperature if necessary
IDS-pH sensor + test sample:
Conductivity too low Use suitable IDS-pH sensor
Temperature too high Use suitable IDS-pH sensor
Organic liquids Use suitable IDS-pH sensor