ba77169e01 04/2016 92
Configuring the
automatic memory
function
Settings You can configure the automatic data storing function with the following set-
tings:
1 Press the <STO_> key.
The menu for automatic data storing appears.
1 Specified entire storing duration
2 Max. available storing duration
3 Graphical display of the memory usage
ID number 1
Interval 30 s
Duration 180 min
continue
0d03h00min
0 1d17h33mi
n
Automatic data stor-
30.09.2015 08:00
1
2
3
Menu item Possible
setting
Explanation
ID number 1 ... 10000 ID number for the dataset series.
Interval 1 s, 5 s,
10 s, 30 s,
1min,
5min,
10 min,
15 min,
30 min,
60 min
Storing interval.
The lower limit of the storing interval can be
restricted by the number of free memory
locations.
The upper limit is restricted by the storing
duration.
Duration 1 min ... x
min
Storing duration.
Specifies after which time the automatic
data storing should be terminated.
The lower limit of the storage duration is
restricted by the storage interval.
The upper limit is restricted by the number
of free memory locations.

Table of Contents