What to do if... OxiTop
®
Control OC 100
70 ba31114e06 06/2012
undef.
(Display in the curve
presentation or sample
statistics)
The selected measurement range
has been exceeded.
The sample filled was too warm.
The AutoTemp function is
switched off.
The sample filled was very cold
and has a low consumption
behavior (small BOD value).
The system is not sealed
(bottle internal pressure = atmo-
spheric pressure).
Select the correct measurement
range (see application reports on
the CD-ROM, Principles of measur-
ing technique).
Pre-temper the sample more pre-
cisely.
Switch on the AutoTemp function
(only effective with run durations of
more than 1 day).
Check the system for leaks by
means of a visual check:
Are the sealing surfaces of the bot-
tle and measuring head clean and
fault-free? Are there cracks in the
sleeve, bottle or measuring head?
Perform a cal-test (see section
7.5.4
CAL-TEST, page 50).
Check the incubator temperature.
10 of 12
called up
(sample message, no. is
variable)
Two measuring heads have not
been found by the controller
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations)
The measuring heads have no
optical contact with the controller:
Distance too great, angular
position imprecise
IR window of the measuring
head averted from the control-
ler
Shading by other objects
The missing measuring heads are
defective.
Modification of the distances or
angle.
Search at other storage locations.
Search for and check measuring
heads (see section 10.3 G
ENERAL
INFORMATION, page 73)
Display message Cause Remedy