OxiTop
®
Control OC 100 What to do if...
ba31114e06 06/2012 71
0 of 19
called up
(sample message, no. is
variable)
No measuring head was found by
the controller.
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations)
The measuring heads have no
optical contact with the
controller.
Distance too great, angular
position imprecise
IR window of the measuring
head averted from the
controller
Optical shading by other
objects
The controller is defective.
Modification of the distances or
angle
Search at other storage locations
Check the controller (see section
10.3 G
ENERAL INFORMATION, page
73)