OxiTop
®
Control OC 100 What to do if...
ba31114e06 06/2012 71
0 of 19
called up
(sample message, no. is
variable)
No measuring head was found by
the controller.
The measuring heads are placed
at another location (e.g. other lev-
els in the incubator, other incuba-
tors, further storage locations)
The measuring heads have no
optical contact with the
controller.
– Distance too great, angular
position imprecise
– IR window of the measuring
head averted from the
controller
– Optical shading by other
objects
The controller is defective.
Modification of the distances or
angle
Search at other storage locations
Check the controller (see section
10.3 G
ENERAL INFORMATION, page
73)
Display message Cause Remedy