EasyManua.ls Logo

Xtreme TX90 6kVA - PC Calibration

Xtreme TX90 6kVA
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Xtreme Power Conversion Corporaon
TX90 6kVA & 10kVA Service Manual
Page 30
Uninterrupble Power Supply
PC Calibraon
Set the communicaon baud rate to 2400. Set “With Transformerin the calibraon seng to “No” as shown
below.
Connect a resistance load to point A and point B
Fill in the values of the voltage and load% measured from the Power Meter into Calibraon Step 1, then
click “ Calibraon1” icon.
If the buzzer beeps intermiently and connuously, it means your calibraon is in failure. You may click
skip step 2” icon and repeat
If the buzzer beeps twice, it means the Calibraon Step 1 is done successfully, then you may enter into
Calibraon Step 2. (The ”ECO”LED light up to indicate the UPS is working under the Calibraon Step2).
Remove the load to “Null” status, then remove the power meter to the output of the terminal block of
the UPS.
Set a correct wiring diagram of the output transformer from the “Seng Calibraon parametermenu
indicated as below.

Related product manuals