INSTRUMENT DESCRIPTION
Cell Observer SD Technical Data Carl Zeiss
M60-2-0021 e 01/2009 423638-7044-001 35
SVB 1 (Signal distribution box)
Line voltage...............................................................................................................100 to 240 V ±10 %
Change of instrument voltage setting is not necessary!
Line frequency .............................................................................................................................50/60 Hz
Maximum power consumption ........................................................................................................ 20 VA
Fuses .......................................................................................................................................2x T1.0 A/H
Intended site..........................................................................................................................closed rooms
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical safety .............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)
under consideration of CSA and UL directives
Overvoltage category............................................................................................................................... II
Radio interference suppression ..........................................................in accordance with EN 55011 class B
Noise immunity................................................................................... in accordance with DIN EN 61326-1
Control Box CSU-X1
Line voltage...............................................................................................................100 to 240 V ±10 %
Change of instrument voltage setting is not necessary!
Line frequency .............................................................................................................................50/60 Hz
Maximum power consumption ..............................................................................................max. 200 VA
Fuses ...................................................................................................................automatic circuit breaker
Intended site..........................................................................................................................closed rooms
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical safety .............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)
Overvoltage category............................................................................................................................... II
Noise immunity................................................................................... in accordance with DIN EN 61326-1
HXP 120
Line voltage...............................................................................................................100 to 240 V ±10 %
Change of instrument voltage setting is not necessary!
Line frequency .............................................................................................................................50/60 Hz
Maximum power consumption ..............................................................................................max. 210 VA
Fuses ......................................................................................................................................2x T 2.5 A/H
Intended site..........................................................................................................................closed rooms
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical safety .............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)