INSTRUMENT DESCRIPTION
Carl Zeiss Technical Data Cell Observer SD
36 423638-7044-001 M60-2-0021 e 01/2009
Definite Focus controller
Line voltage...............................................................................................................100 to 240 V ±10 %
Change of instrument voltage setting is not necessary!
Line frequency............................................................................................................................. 50/60 Hz
Maximum power consumption ........................................................................................................ 60 VA
Fuses ......................................................................................................................................2x T 2.0 A/H
Intended site .........................................................................................................................closed rooms
Protection class......................................................................................................................................... I
Protection type .................................................................................................................................. IP 20
Electrical safety.............................................................in compliance with DIN EN 61010-1 (IEC 61010-1)
under consideration of CSA and UL directives
Overvoltage category............................................................................................................................... II
Radio interference suppression ..........................................................in accordance with EN 55011 class B
Noise immunity .................................................................................. in accordance with DIN EN 61326-1
Power supply for CSU-X1 spinning disc unit
Supply voltage................................................................................... 100 to 240 VAC (± 10 %); 50/60 Hz
Max. power consumption........................................................................................................max. 38 VA
Optical-mechanical data
Scanning field .................................................................................................Standard 10 mm x 7 mm
Mechanical port for microscope .............................................................................................. C-mount
Optical fiber
Fiber type ................................................................................... polarization-preserving mono mode fiber
Transmission.................................................................................................................................. ≥ 50 %
Permissible wavelength range.............................................................................................405 to 650 nm
Beam diameter at inlet opening................................................................................................... 0.65 mm
Type of fiber connection...................................................................................................................... FCP