Service Kit Installation Procedures
906-0731-04-01 Rev. C ZOLL Ventilator Service Manual 4–93
12 Secure the SPM Mix Manifold to the SPM Oxygen
Inlet Manifold. Ensure that the O-Ring does not
move while completing this operation. Failure to
do so will cause the device to fail.
NOTE: SPM Mix Manifold must be flush with the
SPM Oxygen Inlet Manifold.
13 Insert four of Screw, Phillips, Pan Head, 4-40 X 1
¾ (P/N 358-0440-28) into the SPM Oxygen Inlet
Manifold and torque in a crossing pattern to 5 in-
lbs with #1 Phillips head.
Step Action