Pannoramic SCAN II 2.2.0 Users Guide 6.3 Troubleshooting
Slide is often positioned incorrectly
Guiding fault in transport system
Check if faulty positioning disappears after the selection of a different type of
magazine (for example, universal magazine).
Contact 3DHISTECH Service for support.
Digitization process is extremely slow
Extended focusing option is activated
Deactivate the Extended focusing option if not required.
Z-stack option is activated
Deactivate the Z-stack option if not required.
Image is out of focus (in Manual mode)
Focus range limit is activated
Deactivate the Limiting focus range option.
Focus range setting is incorrect
Change focus range setting.
All individual images show a shading effect
Algorithm failed to find empty image fields
Capture a reference image with a blank slide (see chapter 5.6.2.Service - Image
compensation).
6.3.1 Fluorescent Applications (optional)
The switching of the Lumencor SPECTRA 6 cannot be triggered, in spite of trigger connection
through the camera.
Contact 3DHISTECH Service for support.
No suitable exposure time can be found. Signal is too weak.
Lumencor SOLA SM II is switched off
Switch light source on, otherwise contact 3DHISTECH Service for support.
January 24, 2020 - Rev. 2 3DHISTECH Ltd. 137(156)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the 3DHISTECH Pannoramic SCAN II and is the answer not in the manual?

3DHISTECH Pannoramic SCAN II Specifications

General IconGeneral
Brand3DHISTECH
ModelPannoramic SCAN II
CategoryScanner
LanguageEnglish

Related product manuals