Pannoramic SCAN II 2.2.0 Users Guide 5.1.1 Profile Settings Panel
5.1.1 Profile Settings Panel
Objective settings
This field displays the objective used for scanning.
Objective type
20x or 40x
Output resolution
Fix values are: 37x Native for 20x objective, and 73x Native for 40x objective.
Focus settings for the scan profile
Figure 19 – Focus settings for scan profile
Limits of the range: You can define focus limits by selecting Limits of the range for focusing.
You can set the upper and lower limits by selecting the desired value in the selection boxes.
This is useful if focusing was not targeted on the specimen, but on a spot on the coverslip.
Focus distance: the distance of focal points can be set to create a focus map, based on which
the software calculates interpolated focus values (one step in distance means one FOV).
January 24, 2020 - Rev. 2 3DHISTECH Ltd. 54(156)