60 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
8 Operation
Parameter descriptions
Submenu / parameter Value range Remark
Submenu
This menu does not appear for the PROFIBUS PA or
FOUNDATION fieldbus.
Prog. Unit
Unit name Text entry Four character name for the user programmable unit.
kgal = Factory setting.
kgal
Units factor
Numerical entry Reference factor for the user-defined units based on liters;
default value in kgal units (factory setting).
3785.1
Prog. Unit Without density
With density
Unit for mass (with density) or volume (without density).
Without density
Density
Numerical entry Density range 0.1 ... -5 g/cm
3
. Mass flowrate for display and
totalization. The density is only considered when mass flow units
are and/or totalizer mass units are selected.
1 g/cm
3
Submenu
System data
Submenu
Transmitter
Submenu
This menu does not appear with an excitation frequency of 60 Hz
or 70 Hz. Only for old (legacy) sensors.
Data 50 Hz
Channel View only Adjustment value for Channel can be monitored for the
excitation frequency setting (displayed when an older primary 50
Hz has been selected).
-123 µs
Zero
View only
Adjustment value for zero point can be monitored to check the
excitation frequency setting (displayed when an older 50 Hz
sensor has been selected).
0 %
Voltage → V View only Adjustment value for Span Forward can be monitored for the
excitation frequency setting (displayed when an older primary 50
Hz has been selected).
91 %
Voltage ← R
View only
Adjustment value for Span Reverse can be monitored for the
excitation frequency setting (displayed when an older primary 50
Hz has been selected).
-91 %