FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 61
Submenu / parameter Value range Remark
Submenu
This menu does not appear with an excitation frequency of 50 Hz
or 70 Hz. Only for old (legacy) sensors.
Data 60 Hz
Channel
View only Adjustment value for channel can be monitored to check the
excitation frequency setting (displayed when an older 60 Hz
sensor has been selected).
-123 µs
Zero
View only
Adjustment value for zero point can be monitored to check the
excitation frequency setting (displayed when an older 60 Hz
sensor has been selected).
0 %
Voltage → V
View only
Adjustment value for Span Forward can be monitored to check
the excitation frequency setting (displayed when an older 60 Hz
sensor has been selected).
91 %
Voltage ← R
View only
Adjustment value for Span Reverse can be monitored to check
the excitation frequency setting (displayed when an older 60 Hz
sensor has been selected).
-91 %