70 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 8 Operation
… Parameter descriptions
Submenu / parameter Value range Remark
TB VolFlow Value. Displays the value for Volume_Flow
(Transducer block Index 17).
TB VolFlow Status. Displays the status of Volume_Flow
(Transducer block Index 17).
TB Total → F Value. Displays the value for Totalizer → F
(Transducer block Index 102).
TB Total → Status. Displays the status of the Totalizer → F
(Transducer block Index 102).
TB Total ← R Value. Displays the value for Totalizer ← R
(Transducer block Index 104).
TB Total ← R Status. Displays the status of Totalizer ← R
(Transducer block Index 104).
TB Total Diff Value. Displays the value for the diff. totalizer
(Transducer block Index 106).
TB Total Diff Status. Displays the status of the diff. totalizer
(Transducer block Index 106).
Display 1st line FB AI OUT. Displays the OUT value of the AI block. The decimal
places are generated from the OUT_SCALE structure. The unit
displayed is UNIT_INDEX from the OUT_SCALE structure.
Display 1st line FB AI Status. Displays the current mode of the particular block
and the status of the output variables (OUT. Status). The
substatus is displayed after the status, if applicable. Example:
BAD 3 means status is BAD, substatus 3 = device failure.
For information about the numeric code, refer to the interface
documentation.