TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 8
Editing the Calibration Constants
These functions are for recording or modifying slope or offset calibra-
tion constants.
EDIT CONDUCTIVITY
This procedure steps through the modification of the conductivity.
Use the numeric keys to enter the process
temperature.
Key Display Comments
ENTER
CHANGE TEMP FOR CALI-
BRATION
nn.nn °C
ENTER
WORKING
CALIBRATION
COMPLETE
ENTER
ID TAGNAME
READY
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
YES