TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 11
Reset to Factory Configuration
This function resets all of the device parameters to the factory set-
tings. Refer to the product instruction for information on the factory
settings.
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select RESET CAL.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→ YES
ENTER
CALIBRATION
PROCESS CAL
EDIT CAL
→ RESET CAL
ENTER
WORKING
CALIBRATION
COMPLETE
ID TAGNAME
READY