TB82 TRANSMITTER
CALIBRATION
N - 8
Temperature Calibration
This procedure describes how to perform a temperature calibration.
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Selecting YES will hold the
output during calibration.
Select YES.
Select TEMP. CAL.
Enter the temperature. Default temperature is 25°C.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
YES
ENTER
nnnn CAL
TEMP. CAL
EDIT CAL
RESET CAL
ENTER
ENTER
TEMPERATURE:
nn.nn
O
C
ENTER
ID TAGNAME
READY