TB82 TRANSMITTER
CALIBRATION
N - 7
Enter a process temperature value.
This screen appears only for 2 POINT CAL when
performing a pH CAL. 25
°C is the default value.
Enter the low value of the buffer or standard and
press . Enter the high value.
The low default value is 4.0 pH. The high default
value is 7.0 pH.
Apply the low value of the buffer or standard.
Apply the high value of the buffer or standard.
The terminal displays the calculated efficiency in per-
centage.
The terminal displays the calculated sensor offset in
millivolts.
Select YES to release output hold. Screen appears
at the end of two-point calibration when device is in
the analog mode.
If any warnings or messages exist pertaining to this
device, the terminal will list them.
Key Display Comments
ENTER
ENTER PROCESS
TEMPERATURE
nn.nn
O
C
ENTER
LOW CALIB
nn UNITS
HIGH CALIB
nn UNITS
ENTER
ENTER
APPLY n.nn UNITS
PRESS ENTER
WHEN STABLE
ENTER
APPLY n.nn UNITS
PRESS ENTER
WHEN STABLE
ENTER
CALCULATED
EFFICIENCY
nn.n %
PRESS ENTER
ENTER
CALCULATED
SENSOR OFFSET
nnn.n mV
PRESS ENTER
ENTER
RELEASE OUTPUT HOLD
NO
YES
ENTER
ID TAGNAME
READY

Table of Contents