TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
J - 10
Key Display Comments
This operation will cause a change in output not cor-
responding to the input. Be sure the control loop is in
manual.
Select YES.
Select EDIT CAL.
Select TEMPERATURE.
To view values, press twice.
To change the slope, enter the correct number and
press .
To change the offset, enter the correct number and
press .
Calibration complete.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED! PROCEED?
NO
→ YES
ENTER
CALIBRATION
PROCESS CAL
→ EDIT CAL
RESET CAL
ENTER
EDIT CAL DATA
CONDUCTIVITY
→ TEMPERATURE
ENTER
TEMPERATURE
PROCESS CAL
SLP: n.nnnn
OFF:n.nnnn
ENTER
ENTER
ENTER
ENTER
WORKING
CALIBRATION
COMPLETE
ENTER
ID TAGNAME
READY