Suppressed Control Result Display with a Passed QC run: All analyte results will display a <>
symbol and the overall Pass assessment appears at the top of the display screen above the
results.
Suppressed Control Result Display with a Failed QC run: All analyte results will display a <>
symbol and the overall Fail assessment appears at the top of the display screen above the
results. The status of individual failed results is displayed on the handheld as either an up-arrow
(h) for out-of-range high results, or a down-arrow (i) for out-of-range low results.